Welcome!
|
|
Details on ELMBs
35 ATMEGA128L ELMBs are connected to 2 CANbus lines
with total length of approx 330 m. To avoid cross-talk between different
detectors and their electronics, it was decided to have independent ELMBs for
each detectors or sub-system, whenever possible. The distribution of the ELMBs
over the experimental hall is optimized in order to have the shortest wiring
connections to sensors or monitoring points.
The monitoring is done using CANOpen OPC Server (CanOpc.exe), running on
PCCOMPASS08 (Windows XP), installed in DCS room HNB423. The configuration file
for the OPC Server is located in the "OPC" directory. A PVSS API manager
"PVSS00opc -num 7", as a client, connects to the OPCCanOpen Server. 2 ports of a
Kvaser PCI-CAN Q (4 ports) interface card are used (the card is in 4 indepent
ports configuration). The CANbus network consists of 2 CANbus branches, of 120
m and 210 m. Due to their long length (>50 m), cable with 6
conducting wires with cross-section 0.5 mm2 for each is used for CANbus1 and
CANbus2 (multiconductor screened cable layer twisted, CERN catalogue -
04.21.51.106.0).
A special transition board is used for the ELMB connection to the network cable
(see the scheme here). The wires of thick cable are ended by pins and screws on
board contacts, which has printed wiring between to Dsub 9 pin connector. The
ELMB CANbus connector connects to the transition board via a short(0,5 m) flat
cable.
Each ELMB has its own local power supply and there is only one central power
supply for the CANbus. The system has full optocoupling between nodes connected
to CANbus line. The power supply for the CANbus provides 12V and a current of
1.5-2 Amps, being located in the DCS room, close to the PCCOMPASS08. This powers
the CANbus receivers/transmitters, which are optically insulated from the ELMB
internal circuit. The ELMBs have their own local power supplies. As a rule, each
ELMB has a power supply for its analog part and another for the digital part. In
some cases one common power supply is used for both parts. The nominal voltage
of the power supplies is 12V. There are 2 types of power supplies used as local:
Switching power supply OMRON model S82K with 12V output, Imax=1.5A (mounted in a
rail and with protection cover); and Regulated Adapter type EP2, FRIWO with
output 12V, Imax=0.5A.
Plugged to the ELMB's motherboard, 2 types of adapters are used, depending of
the input signals to the ELMB: High performance Pt100 adapters, with Rc = 3.9
KOhm, Rs = 100 Ohm (to be used with Pt4w temperature sensors); and Differential
attenuators with 1:100 divider (100 KOhm/1 KOhm).
Different kinds of sensors are used, depending on the parameters to be
monitored. For temperature measurements we use Pt100 sensors. The humidity
measurements are done using type HIH 3605A humidity sensors. For pressure
measurements, type 144S PCB Series is used. The magnetic field is measured with
UGN35 Hall probe.
In PVSS, there are several monitoring parameters with associated alarm signals
defined (so that an alarm will be displayed if the read value goes out of a
given min/max threshold). The alarm causes changing the background colour of the
alarm text window, from green to orange or red. Also, the alarm signal will be
propagated to the top of the monitoring system hierarchy and cause changing the
colour of the associated detector button on the main COMPASS control panel.
The monitoring parameters that can produce alarm signals are:
-
TARGET, RICH, MM, DC4, STRAWs, W45, MW1 and ECAL1 Temperature Sensors;
-
LV from MW1, MW2, MWPCs, ECAL1 and ECAL2;
-
MW2_gas_alarm. This parameter represents the status of gas system for
MW2 detector;
-
SM1_M_F. Measurement of the stability of SM1 magnetic field - Magnets
panel;
-
B6_M_F. Measurement of the stability of B6 magnetic field - Magnets
panel;
-
SM2 Measurement from the NMR - Magnets panel;
-
Hall pressure - Environment Panel;
DAQ and Tigger room temperatures - Environment panel;
Supply water temperature of raw water circuit (for cooling of Straws, MM and DC) - Environment panel;
Water temperature of tank for water circuit that homegenizes temperature of DC4 frame - DC>DC4>Temperatures
panel;
-
T_W_DEP; T_W_RET. These 2 parameters represent the temperatures of
incoming and outgoing water in the RICH electronics cooling system - Environment
panel.
-
Ecal1_Laser_isOn - ECAL1 panel.
ELMBs configuration
Jump to ELMB #:
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
22
23
24
25
29
30
31
32
33
34
35
- ELMB #1
CANbus line #2
ADC mode: unipolar; ADC range: 100 mV.
Place: Gallerie;
Measurements: LV from RICH Wall.
Status: Installed but not connected to detector.
- ELMB #2
CANbus line #2
ADC mode: unipolar; ADC range: 100 mV.
Place: MW1 FE LV power supply rack, SM2 region, Jura side.
Status: installed
Comments: Cable entering ELMB channels
0-15 labelled "ADC input 1"; to channels 16-31 labelled "ADC input 3";
to channels 32-47 labelled "ADC input 2". In PVSS the factors 0.6, 1.4
and 3.2 in the dp_fct are due to the VME crate itself: 10V means Vout=6V, etc.
ADC channels |
Dp name |
Comments |
Status |
0 |
mW1_C31_5Vp |
Crate #3, +5V, channel 1 |
ON |
1 |
mW1_C31_5Vn |
Crate #3, -5V, channel 1 |
ON |
2 |
mW1_C32_5Vp |
Crate #3, +5V, channel 2 |
ON |
3 |
mW1_C32_5Vn |
Crate #3, -5V, channel 2 |
ON |
4 |
mW1_C41_5Vp |
Crate #4, +5V, channel 1 |
ON |
5 |
mW1_C41_5Vn |
Crate #4, -5V, channel 1 |
ON |
6 |
mW1_C42_5Vp |
Crate #4, +5V, channel 2 |
ON |
7 |
mW1_C42_5Vn |
Crate #4, -5V, channel 2 |
ON |
8 |
mW1_C51_5Vp |
Crate #5, +5V, channel 1 |
ON |
9 |
mW1_C51_5Vn |
Crate #5, -5V, channel 1 |
ON |
11 |
mW1_C52_5Vp |
Crate #5, +5V, channel 2 |
ON |
12 |
mW1_C52_5Vn |
Crate #5, -5V, channel 2 |
ON |
16 |
mW1_V3_5Vp |
Crate VME #3, +5V voltage |
ON |
17 |
mW1_V3_5Ip |
Crate VME #3, +5V current |
ON |
18 |
mW1_V3_12Vp |
Crate VME #3, +12V voltage |
ON |
19 |
mW1_V3_12Ip |
Crate VME #3, +12V current |
ON |
20 |
mW1_V3_12Vn |
Crate VME #3, -12V voltage |
ON |
21 |
mW1_V3_12In |
Crate VME #3, -12V current |
ON |
22 |
mW1_V3_OVT |
Crate VME #3 PS Over-Temperature |
ON |
23 |
mW1_V3_OLD |
Crate VME #3 cards Over-
temperature |
ON |
24 |
mW1_V1_5Vp |
Crate VME #1, +5V voltage |
ON |
33 |
mW1_V1_5Ip |
Crate VME #1, +5V current |
ON |
34 |
mW1_V1_12Vp |
Crate VME #1, +12V voltage |
ON |
35 |
mW1_V1_12Ip |
Crate VME #1, +12V current |
ON |
36 |
mW1_V1_12Vn |
Crate VME #1, -12V voltage |
ON |
37 |
mW1_V1_12In |
Crate VME #1, -12V current |
ON |
38 |
mW1_V1_OVT |
Crate VME #1 PS Over-Temperature |
ON |
39 |
mW1_V1_OLD |
Crate VME #1 cards Over-
temperature |
ON |
40 |
mW1_V2_5Vp |
Crate VME #2, +5V voltage |
ON |
41 |
mW1_V2_5Ip |
Crate VME #2, +5V current |
ON |
42 |
mW1_V2_12Vp |
Crate VME #2, +12V voltage |
ON |
43 |
mW1_V2_12Ip |
Crate VME #2, +12V current |
ON |
44 |
mW1_V2_12Vn |
Crate VME #2, -12V voltage |
ON |
45 |
mW1_V2_12In |
Crate VME #2, -12V current |
ON |
46 |
mW1_V2_OVT |
Crate VME #2 PS Over-Temperature |
ON |
47 |
mW1_V2_OLD |
Crate VME #2 cards Over-
temperature |
ON |
58 - 59 |
T_Mw1_Lv_Crates_Jura |
Pt4w temperature sensor |
ON |
60 - 61 |
T_Mw1_Jura_Floor |
Pt4w temperature sensor, T2' |
ON |
62 - 63 |
pt4w_elmb2_T63 |
Pt4w temperature sensor, T2 |
ON |
- ELMB #3
CANbus line #1 ADC mode: unipolar; ADC range: 100 mV. Place: BMS electronics room. Status: installed.
ADC channels |
Dp name |
Comments |
Status |
32 |
mW2_gas_alarm |
Gas alarm from MW2 gas system |
ON |
36 |
ecal2_hv_signal |
HV system alarm for ECAL2 |
ON |
60 - 61 |
T_W_RET |
Pt4w temp sensor, incoming water for RICH electronics cooling |
ON |
62 - 63 |
T_W_DEP |
Pt4w temp sensor, outgoing water from RICH electronics cooling |
ON |
- ELMB #4
CANbus line #2 ADC mode: unipolar; ADC range: 1 V. Place: MW1 FE LV power supply rack, SM2 region, Saleve side. Status: installed.
Comments: In PVSS the factors 0.6, 1.4
and 3.2 in the dp_fct are due to the VME crate itself: 10V means Vout=6V, etc.
ADC channels |
Dp name |
Comments |
Status |
1 |
mW1_C11_5p |
Crate #1, +5V, channel 1 |
ON |
2 |
mW1_C11_5n |
Crate #1, -5V, channel 1 |
ON |
3 |
mW1_C12_5p |
Crate #1, +5V, channel 2 |
ON |
4 |
mW1_C12_5n |
Crate #1, -5V, channel 2 |
ON |
5 |
mW1_C21_5p |
Crate #2, +5V, channel 1 |
ON |
6 |
mW1_C21_5n |
Crate #2, -5V, channel 1 |
ON |
7 |
mW1_C22_5p |
Crate #2, +5V, channel 2 |
ON |
8 |
mW1_C22_5n |
Crate #2, -5V, channel 2 |
ON |
16 |
mW1_V4_5Vp |
Crate VME #4, +5V voltage |
ON |
17 |
mW1_V4_5Ip |
Crate VME #4, +5V current |
ON |
18 |
mW1_V4_12Vp |
Crate VME #4, +12V voltage |
ON |
19 |
mW1_V4_12Ip |
Crate VME #4, +12V current |
ON |
20 |
mW1_V4_12Vn |
Crate VME #4, -12V voltage |
ON |
21 |
mW1_V4_12In |
Crate VME #4, -12V current |
ON |
22 |
mW1_V4_OVT |
Crate VME #4 PS Over-Temperature |
ON |
23 |
mW1_V4_OLD |
Crate VME #4 cards Over-Temperature |
ON |
24 |
mW1_V5_5Vp |
Crate VME #5, +5V voltage |
ON |
25 |
mW1_V5_5Ip |
Crate VME #5, +5V current |
ON |
26 |
mW1_V5_12Vp |
Crate VME #5, +12V voltage |
ON |
27 |
mW1_V5_12Ip |
Crate VME #5, +12V current |
ON |
28 |
mW1_V5_12Vn |
Crate VME #5, -12V voltage |
ON |
29 |
mW1_V5_12In |
Crate VME #5, -12V current |
ON |
30 |
mW1_V5_OVT |
Crate VME #5 PS Over-Temperature |
ON |
31 |
mW1_V5_OLD |
Crate VME #5 cards Over-Temperature |
ON |
62 - 63 |
T_Mw1_Lv_Crates_Saleve |
Pt4w temperature sensor |
ON |
- ELMB #5
CANbus line #2 ADC mode: unipolar; ADC range: 100 mV. Place: SciFiG FE LV power supply
rack, SM2 region, Saleve side. Status: installed
ADC channels |
Dp name |
Comments |
Status |
60 - 61 |
Mag_Sm1_T_Surface |
Pt4w temperature sensor |
ON |
62 - 63 |
pt4w_elmb5_T63 |
Temperature inside hall 888 (1.5 m away from rack), T5 |
ON |
- ELMB #6
CANbus line #2 ADC mode: unipolar; ADC range: 100mV. Place: STRAW frame, temperature and humidity sensors, Jura side. Status: installed.
ADC channels |
Dp name |
Comments |
Status |
0 - 1 |
STO3V2_TDS |
Pt4w temperature sensor |
ON |
2 - 3 |
STO3V2_TDM |
Pt4w temperature sensor |
ON |
4 - 5 |
STO3V2_TDJ |
Pt4w temperature sensor |
ON |
6 - 7 |
STO3V2_TMS |
Pt4w temperature sensor |
ON |
8 - 9 |
STO3V2_TMJ |
Pt4w temperature sensor |
ON |
10 - 11 |
STO3V2_TUS |
Pt4w temperature sensor |
ON |
12 - 13 |
STO3V2_TUM |
Pt4w temperature sensor |
ON |
14 - 15 |
STO3V2_TUJ |
Pt4w temperature sensor |
ON |
16 - 17 |
STO3Y2_TDS |
Pt4w temperature sensor |
ON |
18 - 19 |
STO3Y2_TDM |
Pt4w temperature sensor |
ON |
20 - 21 |
STO3Y2_TDJ |
Pt4w temperature sensor |
ON |
22 - 23 |
STO3Y2_TMS |
Pt4w temperature sensor |
ON |
24 - 25 |
STO3Y2_TMJ |
Pt4w temperature sensor |
ON |
26 - 27 |
STO3Y2_TUS |
Pt4w temperature sensor |
ON |
28 - 29 |
STO3Y2_TUM |
Pt4w temperature sensor |
ON |
30 - 31 |
STO3Y2_TUJ |
Pt4w temperature sensor |
ON |
32 - 33 |
STO3X2_TDS |
Pt4w temperature sensor |
ON |
34 - 35 |
STO3X2_TDM |
Pt4w temperature sensor |
ON |
36 - 37 |
STO3X2_TDJ |
Pt4w temperature sensor |
ON |
38 - 39 |
STO3X2_TMS |
Pt4w temperature sensor |
ON |
40 - 41 |
STO3X2_TMJ |
Pt4w temperature sensor |
ON |
42 - 43 |
STO3X2_TUS |
Pt4w temperature sensor |
ON |
44 - 45 |
STO3X2_TUM |
Pt4w temperature sensor |
ON |
46 - 47 |
STO3X2_TUJ |
Pt4w temperature sensor |
ON |
48 |
STO3V2_H |
Humidity sensor |
ON |
49 |
STO3X2_H |
Humidity sensor |
ON |
50 |
STO3Y2_H |
Humidity sensor |
ON |
51 |
STO3X1_H |
Humidity sensor |
ON |
52 |
STO3U1_H |
Humidity sensor |
ON |
53 |
STO3Y1_H |
Humidity sensor |
ON |
- ELMB #7
CANbus line #2 ADC mode: unipolar; ADC range: 100 mV. Place: RICH rack; temperature sensors inside RICH vessel, Saleve side, and inside ECal1 cassettes, downstream side, top center and top Jura. Status: installed.
ADC channels |
Dp name |
Comments |
Status |
0 - 1 |
pt4w-RICH-1 |
Pt4w temperature sensor inside vessel |
OFF |
2 - 3 |
pt4w-RICH-2 |
Pt4w temperature sensor inside vessel |
OFF |
4 - 5 |
pt4w-RICH-3 |
Pt4w temperature sensor inside vessel |
ON |
6 - 7 |
pt4w-RICH-4 |
Pt4w temperature sensor inside vessel |
ON |
8 - 9 |
pt4w-RICH-5 |
Pt4w temperature sensor inside vessel |
ON |
10 - 11 |
pt4w-RICH-6 |
Pt4w temperature sensor inside vessel |
ON |
12 - 13 |
pt4w-RICH-7 |
Pt4w temperature sensor inside vessel |
ON |
14 - 15 |
pt4w-RICH-8 |
Pt4w temperature sensor inside vessel |
ON |
16 - 17 |
pt4w-RICH-9 |
Pt4w temperature sensor inside vessel |
ON |
18 - 19 |
pt4w-RICH-10 |
Pt4w temperature sensor inside vessel |
ON |
20 - 21 |
pt4w-RICH-11 |
Pt4w temperature sensor inside vessel |
ON |
22 - 23 |
pt4w-RICH-12 |
Pt4w temperature sensor inside vessel |
ON |
24 - 25 |
pt4w-RICH-13 |
Pt4w temperature sensor inside vessel |
ON |
26 - 27 |
pt4w-RICH-14 |
Pt4w temperature sensor inside vessel |
ON |
28 - 29 |
pt4w-RICH-15 |
Pt4w temperature sensor inside vessel |
ON |
30 - 31 |
pt4w-RICH-16 |
Pt4w temperature sensor inside vessel |
ON |
32 - 33 |
pt4w-RICH-17 |
Pt4w temperature sensor inside vessel |
ON |
34 - 35 |
pt4w-RICH-18 |
Pt4w temperature sensor inside vessel |
ON |
48 - 49 |
PT_4W_48_49 |
Pt4w temperature sensor in ECal1 downstream side, Gams Saleve middle |
ON |
50 - 51 |
PT_4W_50_51 |
Pt4w temperature sensor in ECal1 downstream side, Mainz Saleve top |
ON |
52 - 53 |
PT_4W_52_53 |
Pt4w temperature sensor in ECal1 downstream side, Olga Jura top |
ON |
54 - 55 |
PT_4W_54_55 |
Pt4w temperature sensor in ECal1 downstream side, Olga Jura middle |
ON |
62 - 63 |
pt4w-elmb7-T63 |
Pt4w sensor: temperature inside hall 888, T7 |
ON |
- ELMB #8
CANbus line #2 ADC mode: bipolar; ADC range: 100 mV. Place: SciFi rack, Saleve side. Status: installed.
ADC channels |
Dp name |
Comments |
Status |
0 - 1 |
PT_1_HSO |
Pt4w temp sensor, Pol Target horizontal screen outlet |
ON |
2 - 3 |
PT_2_DTF |
Pt4w temp sensor, Pol Target dilution cryostat top flange |
ON |
4 - 5 |
PT_3_SLT |
Pt4w temp sensor, Pol Target still line top |
ON |
6 - 7 |
PT_4_MHG |
Pt4w temp sensor, Pol Target magnet gas outlet |
OFF |
8 - 9 |
PT_5_SCW |
Pt4w temp sensor, Pol Target magnet downstream neck |
ON |
10 - 11 |
PT_6_SFC |
Pt4w temp sensor, Pol Target separator flow control |
ON |
12 - 13 |
PT_7_NMR |
Pt4w temp sensor, Pol Target NMR rack |
ON |
14 - 15 |
PT_8_SCL |
Pt4w temp sensor, Pol Target magnet upstream neck |
ON |
16 |
SM1_M_F |
Hall probe for SM1 Magnetic field |
ON |
28 - 29 |
PT_4W_28_29 |
Pt4w temp sensor, LH2 target gas control panel |
OFF |
30 - 31 |
PT_4W_30_31 |
Pt4w temp sensor, LH2 target hidrogen tank |
OFF |
32 - 33 |
SI01_T_u1 |
Pt4w temp sensor |
OFF |
34 - 35 |
SI01_T_u2 |
Pt4w temp sensor |
ON |
36 - 37 |
SI02_T_u1 |
Pt4w temp sensor |
ON |
38 - 39 |
SI02_T_u2 |
Pt4w temp sensor |
ON |
40 - 41 |
SI02_T_y1 |
Pt4w temp sensor |
ON |
42 - 43 |
SI02_T_y2 |
Pt4w temp sensor |
ON |
44 - 45 |
SI01_T_x1 |
Pt4w temp sensor |
ON |
46 - 47 |
SI01_T_x2 |
Pt4w temp sensor |
ON |
48 - 49 |
SI03_T_u1 |
Pt4w temp sensor |
ON |
50 - 51 |
SI03_T_u2 |
Pt4w temp sensor |
ON |
52 - 53 |
SI02_T_v1 |
Pt4w temp sensor |
OFF |
54 - 55 |
SI02_T_v2 |
Pt4w temp sensor |
OFF |
56 - 57 |
SI03_T_y1 |
Pt4w temp sensor |
ON |
58 - 59 |
SI03_T_y2 |
Pt4w temp sensor |
ON |
60 - 61 |
SI02_T_x1 |
Pt4w temp sensor |
ON |
62 - 63 |
pt4w_elmb8_T63 |
Pt4w temperature
sensor inside hall 888, T8 |
ON |
- ELMB #9
CANbus line #1 ADC mode: unipolar; ADC range: 100 mV. Place: BMS HV room HNB 202. Status: installed.
ADC channels |
Dp name |
Comments |
Status |
0 |
B6_M_F |
sensor on Bending Magnet B6 |
ON |
16 - 17 |
PT_4W_16_17 |
Pt4w sensor, CEDAR1 tent temperature |
ON |
18 - 19 |
PT_4W_19_19 |
Pt4w sensor, CEDAR1 tent temperature |
ON |
20 - 21 |
PT_4W_20_21 |
Pt4w sensor, CEDAR1 tent temperature |
ON |
22 - 23 |
PT_4W_22_23 |
Pt4w sensor, CEDAR1 tent temperature |
ON |
24 - 25 |
PT_4W_24_25 |
Pt4w sensor, CEDAR1 tent temperature |
ON |
26 - 27 |
PT_4W_26_27 |
Pt4w sensor, CEDAR1 tent temperature |
ON |
28 - 29 |
PT_4W_28_29 |
Pt4w sensor, CEDAR1 tent temperature |
ON |
30 - 31 |
PT_4W_30_31 |
Pt4w sensor, CEDAR1 tent temperature |
ON |
32 - 33 |
PT_4W_32_33 |
Pt4w sensor, CEDAR2 tent temperature |
ON |
34 - 35 |
PT_4W_34_35 |
Pt4w sensor, CEDAR2 tent temperature |
ON |
36 - 37 |
PT_4W_36_37 |
Pt4w sensor, CEDAR2 tent temperature |
ON |
38 - 39 |
PT_4W_38_39 |
Pt4w sensor, CEDAR2 tent temperature |
ON |
40 - 41 |
PT_4W_40_41 |
Pt4w sensor, CEDAR2 tent temperature |
ON |
42 - 43 |
PT_4W_42_43 |
Pt4w sensor, CEDAR2 tent temperature |
ON |
44 - 45 |
PT_4W_44_45 |
Pt4w sensor, CEDAR2 tent temperature |
ON |
46 - 47 |
PT_4W_66_47 |
Pt4w sensor, CEDAR2 tent temperature |
ON |
60 - 61 |
AirTemp |
Pt4w sensor, Air temperature outside building 888 |
ON |
62 - 63 |
T_OUT_SIDE |
Pt4w sensor, temperature inside BMS room |
ON |
- ELMB #10
CANbus line #2 ADC mode: unipolar; ADC range: 100 mV. Place: bld. 888, mMegas/Drifts electronics rack Status: installed
ADC channels |
Dp name |
Comments |
Status |
0 - 1 |
MM03_UV_1 |
Pt4w temperature sensor for MM |
ON |
2 - 3 |
MM03_UV_2 |
Pt4w temperature sensor for MM |
ON |
4 - 5 |
MM03_UV_3 |
Pt4w temperature sensor for MM |
ON |
6 - 7 |
MM03_UV_4 |
Pt4w temperature sensor for MM |
ON |
8 - 9 |
MM03_XY_1 |
Pt4w temperature sensor for MM |
ON |
10 - 11 |
MM03_XY_2 |
Pt4w temperature sensor for MM |
ON |
12 - 13 |
MM03_XY_3 |
Pt4w temperature sensor for MM |
ON |
14 - 15 |
MM03_XY_4 |
Pt4w temperature sensor for MM |
ON |
16 - 17 |
MM01_UV_1 |
Pt4w temperature sensor for MM |
ON |
18 - 19 |
MM01_UV_2 |
Pt4w temperature sensor for MM |
ON |
20 - 21 |
MM01_UV_3 |
Pt4w temperature sensor for MM |
OFF |
22 - 23 |
MM01_UV_4 |
Pt4w temperature sensor for MM |
ON |
24 - 25 |
MM01_XY_1 |
Pt4w temperature sensor for MM |
OFF |
26 - 27 |
MM01_XY_2 |
Pt4w temperature sensor for MM |
OFF |
28 - 29 |
MM01_XY_3 |
Pt4w temperature sensor for MM |
OFF |
30 - 31 |
MM01_XY_4 |
Pt4w temperature sensor for MM |
OFF |
32 - 33 |
MM02_XY_1 |
Pt4w temperature sensor for MM |
OFF |
34 - 35 |
MM02_XY_2 |
Pt4w temperature sensor for MM |
OFF |
36 - 37 |
MM02_UV_3 |
Pt4w temperature sensor for MM |
ON |
38 - 39 |
MM02_UV_4 |
Pt4w temperature sensor for MM |
OFF |
40 - 41 |
MM02_XY_1 |
Pt4w temperature sensor for MM |
ON |
42 - 43 |
MM02_XY_2 |
Pt4w temperature sensor for MM |
ON |
44 - 45 |
MM02_XY_3 |
Pt4w temperature sensor for MM |
ON |
46 - 47 |
MM02_XY_4 |
Pt4w temperature sensor for MM |
ON |
58 |
Hum1_HALL |
Humidity sensor |
ON |
69 |
Patm_hall |
Atmospheric pressure sensor |
ON |
60 - 61 |
Mag_Sm1_T_Surface |
Pt4w temperature sensor |
ON |
62 - 63 |
pt4w-elmb10-T63 |
Temperature sensor, point T10 |
ON |
- ELMB #11
CANbus line #1 ADC mode: unipolar; ADC range: 100 mV. Place: bld. 888, W45 rack
Comments: Stations 7 and 11 are not connected
since the power supplies have moved from MWPC to W45 rack. Status:
installed
ADC channels |
Dp name |
Comments |
Status |
0 |
mwpc_8X_6v_vmon |
Low voltage for MWPC station 8 |
ON |
1 |
mwpc_8X_6v_imon |
Low voltage for MWPC station 8 |
ON |
2 |
mwpc_8X_9v_vmon |
Low voltage for MWPC station 8 |
ON |
3 |
mwpc_8X_9v_imon |
Low voltage for MWPC station 8 |
ON |
4 |
mwpc_7X_6v_vmon |
Low voltage for MWPC station 7 |
OFF |
5 |
mwpc_7X_6v_imon |
Low voltage for MWPC station 7 |
OFF |
6 |
mwpc_7X_9v_vmon |
Low voltage for MWPC station 7 |
OFF |
7 |
mwpc_7X_9v_imon |
Low voltage for MWPC station 7 |
OFF |
8 |
mwpc_7V_6v_vmon |
Low voltage for MWPC station 7 |
OFF |
9 |
mwpc_7V_6v_imon |
Low voltage for MWPC station 7 |
OFF |
10 |
mwpc_7V_9v_vmon |
Low voltage for MWPC station 7 |
OFF |
11 |
mwpc_7V_9v_imon |
Low voltage for MWPC station 7 |
OFF |
12 |
mwpc_7U_6v_vmon |
Low voltage for MWPC station 7 |
OFF |
13 |
mwpc_7U_6v_imon |
Low voltage for MWPC station 7 |
OFF |
14 |
mwpc_7U_9v_vmon |
Low voltage for MWPC station 7 |
OFF |
15 |
mwpc_7U_9v_imon |
Low voltage for MWPC station 7 |
OFF |
16 |
mwpc_11U_6v_vmon |
Low voltage for MWPC station 11 |
OFF |
17 |
mwpc_11U_6v_imon |
Low voltage for MWPC station 11 |
OFF |
18 |
mwpc_11U_9v_vmon |
Low voltage for MWPC station 11 |
OFF |
19 |
mwpc_11U_9v_imon |
Low voltage for MWPC station 11 |
OFF |
20 |
mwpc_11V_6v_vmon |
Low voltage for MWPC station 11 |
OFF |
21 |
mwpc_11V_6v_imon |
Low voltage for MWPC station 11 |
OFF |
22 |
mwpc_11V_9v_vmon |
Low voltage for MWPC station 11 |
OFF |
23 |
mwpc_11V_9v_imon |
Low voltage for MWPC station 11 |
OFF |
24 |
mwpc_11X_6v_vmon |
Low voltage for MWPC station 11 |
OFF |
25 |
mwpc_11X_6v_imon |
Low voltage for MWPC station 11 |
OFF |
26 |
mwpc_11X_9v_vmon |
Low voltage for MWPC station 11 |
OFF |
27 |
mwpc_11X_9v_imon |
Low voltage for MWPC station 11 |
OFF |
62 - 63 |
pt4w-elmb11-T63 |
Temperature sensor, T11 |
ON |
- ELMB #12
CANbus
line #1 ADC mode: unipolar; ADC range: 100 mV. Place: bld. 888, MWPC power supply rack Status: installed
Comments: ELMB #11 and ELMB #12 are mounted
in one box and have common power supply and external wiring for all CANbus signals.
ADC channels |
Dp name |
Comments |
Status |
0 |
mwpc_10V_6v_vmon |
Low voltage for MWPC station 10 |
ON |
1 |
mwpc_10V_6v_imon |
Low voltage for MWPC station 10 |
ON |
2 |
mwpc_10V_9v_vmon |
Low voltage for MWPC station 10 |
ON |
3 |
mwpc_10V_9v_imon |
Low voltage for MWPC station 10 |
ON |
4 |
mwpc_10U_6v_vmon |
Low voltage for MWPC station 10 |
ON |
5 |
mwpc_10U_6v_imon |
Low voltage for MWPC station 10 |
ON |
6 |
mwpc_10U_9v_vmon |
Low voltage for MWPC station 10 |
ON |
7 |
mwpc_10U_9v_imon |
Low voltage for MWPC station 10 |
ON |
8 |
mwpc_10X_6v_vmon |
Low voltage for MWPC station 10 |
ON |
9 |
mwpc_10X_6v_imon |
Low voltage for MWPC station 10 |
ON |
10 |
mwpc_10X_9v_vmon |
Low voltage for MWPC station 10 |
ON |
11 |
mwpc_10X_9v_imon |
Low voltage for MWPC station 10 |
ON |
12 |
mwpc_9X_6v_vmon |
Low voltage for MWPC station 9 |
ON |
13 |
mwpc_9X_6v_imon |
Low voltage for MWPC station 9 |
ON |
14 |
mwpc_9X_9v_vmon |
Low voltage for MWPC station 9 |
ON |
15 |
mwpc_9X_9v_imon |
Low voltage for MWPC station 9 |
ON |
16 |
mwpc_9V_6v_vmon |
Low voltage for MWPC station 9 |
ON |
17 |
mwpc_9V_6v_imon |
Low voltage for MWPC station 9 |
ON |
18 |
mwpc_9V_9v_vmon |
Low voltage for MWPC station 9 |
ON |
19 |
mwpc_9V_9v_imon |
Low voltage for MWPC station 9 |
ON |
20 |
mwpc_9U_6v_vmon |
Low voltage for MWPC station 9 |
ON |
21 |
mwpc_9U_6v_imon |
Low voltage for MWPC station 9 |
ON |
22 |
mwpc_9U_9v_vmon |
Low voltage for MWPC station 9 |
ON |
23 |
mwpc_9U_9v_imon |
Low voltage for MWPC station 9 |
ON |
24 |
mwpc_8V_6v_vmon |
Low voltage for MWPC station 8 |
ON |
25 |
mwpc_8V_6v_imon |
Low voltage for MWPC station 8 |
ON |
26 |
mwpc_8V_9v_vmon |
Low voltage for MWPC station 8 |
ON |
27 |
mwpc_8V_9v_imon |
Low voltage for MWPC station 8 |
ON |
28 |
mwpc_8U_6v_vmon |
Low voltage for MWPC station 8 |
ON |
29 |
mwpc_8U_6v_imon |
Low voltage for MWPC station 8 |
ON |
30 |
mwpc_8U_9v_vmon |
Low voltage for MWPC station 8 |
ON |
31 |
mwpc_8U_9v_imon |
Low voltage for MWPC station 8 |
ON |
- ELMB #13
CANbus line #2 ADC mode: unipolar; ADC range: 100 mV. Place: bld. 888, MWPC power supply rack,
Jura side. Status: installed
Comments: ELMB #13 and ELMB #14 are mounted
in one box and have common power supply and external wiring for all
CANbus signals.
ADC channels |
Dp name |
Comments |
Status |
0 |
mwpc_6X_6v_vmon |
Low voltage for MWPC station 6 |
ON |
1 |
mwpc_6X_6v_imon |
Low voltage for MWPC station 6 |
ON |
2 |
mwpc_6X_9v_vmon |
Low voltage for MWPC station 6 |
ON |
3 |
mwpc_6X_9v_imon |
Low voltage for MWPC station 6 |
ON |
4 |
mwpc_5V_6v_vmon |
Low voltage for MWPC station 5 |
ON |
5 |
mwpc_5V_6v_imon |
Low voltage for MWPC station 5 |
ON |
6 |
mwpc_5V_9v_vmon |
Low voltage for MWPC station 5 |
ON |
7 |
mwpc_5V_9v_imon |
Low voltage for MWPC station 5 |
ON |
8 |
mwpc_2U_6v_vmon |
Low voltage for MWPC station 2 |
ON |
9 |
mwpc_2U_6v_imon |
Low voltage for MWPC station 2 |
ON |
10 |
mwpc_2U_9v_vmon |
Low voltage for MWPC station 2 |
ON |
11 |
mwpc_2U_9v_imon |
Low voltage for MWPC station 2 |
ON |
12 |
mwpc_2V_6v_vmon |
Low voltage for MWPC station 2 |
ON |
13 |
mwpc_2V_6v_imon |
Low voltage for MWPC station 2 |
ON |
14 |
mwpc_2V_9v_vmon |
Low voltage for MWPC station 2 |
ON |
15 |
mwpc_2V_9v_imon |
Low voltage for MWPC station 2 |
ON |
16 |
mwpc_2X_6v_vmon |
Low voltage for MWPC station 2 |
ON |
17 |
mwpc_2X_6v_imon |
Low voltage for MWPC station 2 |
ON |
18 |
mwpc_2X_9v_vmon |
Low voltage for MWPC station 2 |
ON |
19 |
mwpc_2X_9v_imon |
Low voltage for MWPC station 2 |
ON |
20 |
mwpc_3X_6v_vmon |
Low voltage for MWPC station 3 |
ON |
21 |
mwpc_3X_6v_imon |
Low voltage for MWPC station 3 |
ON |
22 |
mwpc_3X_9v_vmon |
Low voltage for MWPC station 3 |
ON |
23 |
mwpc_3X_9v_imon |
Low voltage for MWPC station 3 |
ON |
24 |
mwpc_3U_6v_vmon |
Low voltage for MWPC station 3 |
ON |
25 |
mwpc_3U_6v_imon |
Low voltage for MWPC station 3 |
ON |
26 |
mwpc_3U_9v_vmon |
Low voltage for MWPC station 3 |
ON |
27 |
mwpc_3U_9v_imon |
Low voltage for MWPC station 3 |
ON |
28 |
mwpc_3V_6v_vmon |
Low voltage for MWPC station 3 |
ON |
29 |
mwpc_3V_6v_imon |
Low voltage for MWPC station 3 |
ON |
30 |
mwpc_3V_9v_vmon |
Low voltage for MWPC station 3 |
ON |
31 |
mwpc_3V_9v_imon |
Low voltage for MWPC station 3 |
ON |
62 - 63 |
Mag_Sm2_T_Water_Depart |
Pt4w temperature sensor |
ON |
- ELMB #14
CANbus line #2 ADC mode: unipolar; ADC range: 100 mV. Place: bld. 888, MWPC power supply rack,
Jura side Status: installed.
Comments: ELMB #13 and ELMB #14 are mounted
in one box and have common power supply and external wiring for all
CANbus signals.
ADC channels |
Dp name |
Comments |
Status |
0 |
mwpc_4U_6v_vmon |
Low voltage for MWPC station 4 |
ON |
1 |
mwpc_4U_6v_imon |
Low voltage for MWPC station 4 |
ON |
2 |
mwpc_4U_9v_vmon |
Low voltage for MWPC station 4 |
ON |
3 |
mwpc_4U_9v_imon |
Low voltage for MWPC station 4 |
ON |
4 |
mwpc_4V_6v_vmon |
Low voltage for MWPC station 4 |
ON |
5 |
mwpc_4V_6v_imon |
Low voltage for MWPC station 4 |
ON |
6 |
mwpc_4V_9v_vmon |
Low voltage for MWPC station 4 |
ON |
7 |
mwpc_4V_9v_imon |
Low voltage for MWPC station 4 |
ON |
8 |
mwpc_4X_6v_vmon |
Low voltage for MWPC station 4 |
ON |
9 |
mwpc_4X_6v_imon |
Low voltage for MWPC station 4 |
ON |
10 |
mwpc_4X_9v_vmon |
Low voltage for MWPC station 4 |
ON |
11 |
mwpc_4X_9v_imon |
Low voltage for MWPC station 4 |
ON |
12 |
mwpc_5X_6v_vmon |
Low voltage for MWPC station 5 |
ON |
13 |
mwpc_5X_6v_imon |
Low voltage for MWPC station 5 |
ON |
14 |
mwpc_5X_9v_vmon |
Low voltage for MWPC station 5 |
ON |
15 |
mwpc_5X_9v_imon |
Low voltage for MWPC station 5 |
ON |
16 |
mwpc_5U_6v_vmon |
Low voltage for MWPC station 5 |
ON |
17 |
mwpc_5U_6v_imon |
Low voltage for MWPC station 5 |
ON |
18 |
mwpc_5U_9v_vmon |
Low voltage for MWPC station 5 |
ON |
19 |
mwpc_5U_9v_imon |
Low voltage for MWPC station 5 |
ON |
21 |
mwpc_6V_6v_vmon |
Low voltage for MWPC station 6 |
ON |
25 |
mwpc_6V_6v_imon |
Low voltage for MWPC station 6 |
ON |
26 |
mwpc_6V_9v_vmon |
Low voltage for MWPC station 6 |
ON |
27 |
mwpc_6V_9v_imon |
Low voltage for MWPC station 6 |
ON |
28 |
mwpc_6U_6v_vmon |
Low voltage for MWPC station 6 |
ON |
29 |
mwpc_6U_6v_imon |
Low voltage for MWPC station 6 |
ON |
30 |
mwpc_6U_9v_vmon |
Low voltage for MWPC station 6 |
ON |
31 |
mwpc_6U_9v_imon |
Low voltage for MWPC station 6 |
ON |
60 - 61 |
Mag_Sm2_T_Water_Return |
Pt4w temperature sensor |
ON |
62 - 63 |
pt4w_elmb14_T63 |
Pt4w temperature sensor inside hall 888, T14 |
ON |
- ELMB #15
CANbus line #2 ADC mode:unipolar; ADC range: 100 mV. Place: bld. 888, on gallery, Jura side Status: installed.
ADC channels |
Dp name |
Comments |
Status |
1 |
mwpc_1X_6v_vmon |
Low voltage for MWPC station 1 |
ON |
2 |
mwpc_1X_6v_imon |
Low voltage for MWPC station 1 |
ON |
3 |
mwpc_1X_9v_vmon |
Low voltage for MWPC station 1 |
ON |
4 |
mwpc_1X_9v_imon |
Low voltage for MWPC station 1 |
ON |
5 |
mwpc_1Y_6v_vmon |
Low voltage for MWPC station 1 |
ON |
6 |
mwpc_1Y_6v_imon |
Low voltage for MWPC station 1 |
ON |
7 |
mwpc_1Y_9v_vmon |
Low voltage for MWPC station 1 |
ON |
8 |
mwpc_1Y_9v_imon |
Low voltage for MWPC station 1 |
ON |
9 |
mwpc_1V_6v_vmon |
Low voltage for MWPC station 1 |
ON |
10 |
mwpc_1V_6v_imon |
Low voltage for MWPC station 1 |
ON |
11 |
mwpc_1V_9v_vmon |
Low voltage for MWPC station 1 |
ON |
12 |
mwpc_1V_9v_imon |
Low voltage for MWPC station 1 |
ON |
13 |
mwpc_1U_6v_vmon |
Low voltage for MWPC station 1 |
ON |
14 |
mwpc_1U_6v_imon |
Low voltage for MWPC station 1 |
ON |
15 |
mwpc_1U_9v_vmon |
Low voltage for MWPC station 1 |
ON |
16 |
mwpc_1U_9v_imon |
Low voltage for MWPC station 1 |
ON |
62 - 63 |
pt4w-elmb15-T63 |
Pt4w temperature sensor inside hall 888, T10 |
OFF |
- ELMB #16
CANbus line #1 ADC mode: bipolar; ADC range: 1 V. Place: Trigger room HNB422 Status: installed.
Comments: Automated system for Gas distribution in the area
provides 3 switches. A small current is injected to give the signal (closed=on=alarm). Also
PS for Veto Box gives current and voltage (scaling factors applied in PVSS according to the PS info
-> 1V=0.4A and 1V=3.33 KV).
ADC channels |
Dp name |
Comments |
Status |
0 - 1 |
Temp_NIM_1 |
Pt4w temperature sensor, inside ELMB box |
OFF |
58 |
VBX_PS_V |
Voltage from Veto Box PS unit |
OFF |
59 |
VBX_PS_I |
Current from Veto Box PS unit |
OFF |
62 |
Gas_lv3_alarm |
Level 3 alarm from Gas gen. distrib. system |
OFF |
63 |
Gas_gen_alarm |
General alarm from Gas gen. distrib. system |
OFF |
64 |
Gas_lv2_alarm |
Level 2 alarm from Gas gen. distrib. system |
OFF |
- ELMB #17
CANbus line #2 ADC mode: unipolar; ADC range: 100 mV. Place: bld. 888, STRAW frame, close to ELMB
#6 Status: spare, NO Data Points in PVSS
- ELMB
#18
CANbus line #1 ADC mode: bipolar; ADC
range: 1 V. Place: Trigger room HNB429 Status: installed.
ADC channels |
Dp name |
Comments |
Status |
16 - 17 |
PT_4W_16_17 |
Pt4w temperature sensor inside trigger room |
ON |
18 - 19 |
PT_4W_18_19 |
Pt4w temperature sensor inside DAQ room |
ON |
25 |
Ecal1_Hv_Laser |
TTL signal for monitoring status of Laser for ECAL1 |
ON |
26 |
Ecal1_Lv_Gams_Voltage |
Low voltage of ECAL1 (GAMS), "100 V" |
ON |
27 |
Ecal1_Lv_Gams_Current |
Low voltage of ECAL1 (GAMS), "100 V" |
ON |
28 |
Ecal1_Lv_Mainz_Olga_Voltage |
Low voltage of ECAL1 (Mainz and OLGA), "100 V" |
ON |
29 |
Ecal1_Lv_Mainz_Olga_Current |
Low voltage of ECAL1 (Mainz and OLGA), "100 V" |
ON |
30 |
Ecal2_Lv_Voltage |
Low voltage of ECAL2, "100 V" |
ON |
31 |
Ecal2_Lv_Current |
Low voltage of ECAL2, "100 V" |
ON |
49 |
Humidity_sensor_49 |
Humidity sensor inside trigger room |
ON |
50 |
Humidity_sensor_50 |
Humidity sensor inside DAQ room |
ON |
51 |
Humidity_sensor_51 |
Humidity sensor inside Veto room |
ON |
62-63 |
PT_4W_62_63 |
Pt4w temperature sensor inside Veto room |
ON |
- ELMB #19
CANbus line #1 ADC mode: unipolar; ADC range 100 mV. Place: bld. 888, beam area downstream, MW2 power supply rack. Status: installed.
ADC channels |
Dp name |
Comments |
Status |
0 |
Ampl X1-1 |
Amplifier for X1, mW2 low voltage PS |
ON |
1 |
Ampl X1-2 |
Amplifier for X1, mW2 low voltage PS |
ON |
2 |
Ampl V1-1 |
Amplifier for V1, mW2 low voltage PS |
ON |
3 |
Ampl V1-2 |
Amplifier for V1, mW2 low voltage PS |
ON |
4 |
Ampl X2-1 |
Amplifier for X2, mW2 low voltage PS |
ON |
5 |
Ampl X2-2 |
Amplifier for X2, mW2 low voltage PS |
ON |
6 |
Ampl V2-1 |
Amplifier for V2, mW2 low voltage PS |
ON |
7 |
Ampl V2-2 |
Amplifier for V2, mW2 low voltage PS |
ON |
9 |
Ampl Y1 |
mW2 low voltage PS |
ON |
11 |
Ampl Y2 |
mW2 low voltage PS |
ON |
12 |
TDC |
mW2 low voltage PS |
ON |
13 |
|
|
OFF |
14 |
LAX2_N2 |
TDC for X,V, mW2 low voltage PS |
ON |
15 |
|
Thresholds Y1, Y2, mW2 low voltage PS |
ON |
16 |
|
Thresholds X1, V1, mW2 low voltage PS |
ON |
17 |
|
Thresholds X2, V2, mW2 low voltage PS |
ON |
62 - 63 |
pt4w_elmb19_T63 |
Pt4w temperature sensor, T19 |
ON |
- ELMB #20
CANbus line #1 ADC mode: unipolar; ADC range: 100 mV. Place: bld. 888, W45 power supply rack. Status: installed.
ADC channels |
Dp name |
Comments |
Status |
0 |
W05_1_6V |
voltage, low voltage PS for W45 |
ON |
1 |
W05_1_6A |
current, low voltage PS for W45 |
ON |
2 |
W05_1_9V |
voltage, low voltage PS for W45 |
ON |
3 |
W05_1_9A |
current, low voltage PS for W45 |
ON |
4 |
W05_2_6V |
voltage, low voltage PS for W45 |
ON |
5 |
W05_2_6A |
current, low voltage PS for W45 |
ON |
6 |
W05_2_9V |
voltage, low voltage PS for W45 |
ON |
7 |
W05_2_9A |
current, low voltage PS for W45 |
ON |
8 |
W04_1_6V |
voltage, low voltage PS for W45 |
ON |
9 |
W04_1_6A |
current, low voltage PS for W45 |
ON |
10 |
W04_1_9V |
voltage, low voltage PS for W45 |
ON |
11 |
W04_1_9A |
current, low voltage PS for W45 |
ON |
12 |
W04_2_6V |
voltage, low voltage PS for W45 |
ON |
13 |
W04_2_6A |
current, low voltage PS for W45 |
ON |
14 |
W04_2_9V |
voltage, low voltage PS for W45 |
ON |
15 |
W04_2_9A |
current, low voltage PS for W45 |
ON |
16 |
W02_1_6V |
voltage, low voltage PS for W45 |
ON |
17 |
W02_1_6A |
current, low voltage PS for W45 |
ON |
18 |
W02_1_9V |
voltage, low voltage PS for W45 |
ON |
19 |
W02_1_9A |
current, low voltage PS for W45 |
ON |
20 |
W02_2_6V |
voltage, low voltage PS for W45 |
ON |
21 |
W02_2_6A |
current, low voltage PS for W45 |
ON |
22 |
W02_2_9V |
voltage, low voltage PS for W45 |
ON |
23 |
W02_2_9A |
current, low voltage PS for W45 |
ON |
24 |
W01_1_6V |
voltage, low voltage PS for W45 |
ON |
25 |
W01_1_6A |
current, low voltage PS for W45 |
ON |
26 |
W01_1_9V |
voltage, low voltage PS for W45 |
ON |
27 |
W01_1_9A |
current, low voltage PS for W45 |
ON |
28 |
W01_2_6V |
voltage, low voltage PS for W45 |
ON |
29 |
W01_2_6A |
current, low voltage PS for W45 |
ON |
30 |
W01_2_9V |
voltage, low voltage PS for W45 |
ON |
31 |
W01_2_9A |
current, low voltage PS for W45 |
ON |
- ELMB #22
CANbus line #2 ADC mode: unipolar; ADC range: 100 mV. Place: bld. 888, Straw station
5 frame. Status: installed.
ADC channels |
Dp name |
Comments |
Status |
0 - 1 |
ST05V1_TDS |
Pt4w temp sensor, Straw station 5 |
ON |
2 - 3 |
ST05V1_TDM |
Pt4w temp sensor, Straw station 5 |
ON |
4 - 5 |
ST05V1_TDJ |
Pt4w temp sensor, Straw station 5 |
ON |
6 - 7 |
ST05V1_TMS |
Pt4w temp sensor, Straw station 5 |
ON |
8 - 9 |
ST05V1_TMJ |
Pt4w temp sensor, Straw station 5 |
ON |
10 - 11 |
ST05V1_TUS |
Pt4w temp sensor, Straw station 5 |
ON |
12 - 13 |
ST05V1_TUM |
Pt4w temp sensor, Straw station 5 |
ON |
14 - 15 |
ST05V1_TUJ |
Pt4w temp sensor, Straw station 5 |
ON |
16 - 17 |
ST05Y1_TDS |
Pt4w temp sensor, Straw station 5 |
ON |
18 - 19 |
ST05Y1_TDM |
Pt4w temp sensor, Straw station 5 |
ON |
20 - 21 |
ST05Y1_TDJ |
Pt4w temp sensor, Straw station 5 |
ON |
22 - 23 |
ST05Y1_TMS |
Pt4w temp sensor, Straw station 5 |
ON |
24 - 25 |
ST05Y1_TMJ |
Pt4w temp sensor, Straw station 5 |
ON |
26 - 27 |
ST05Y1_TUS |
Pt4w temp sensor, Straw station 5 |
ON |
28 - 29 |
ST05Y1_TUM |
Pt4w temp sensor, Straw station 5 |
ON |
30 - 31 |
ST05Y1_TUJ |
Pt4w temp sensor, Straw station 5 |
ON |
32 - 33 |
ST05X1_TDS |
Pt4w temp sensor, Straw station 5 |
ON |
34 - 35 |
ST05X1_TDM |
Pt4w temp sensor, Straw station 5 |
ON |
36 - 37 |
ST05X1_TDJ |
Pt4w temp sensor, Straw station 5 |
ON |
38 - 39 |
ST05X1_TMS |
Pt4w temp sensor, Straw station 5 |
ON |
40 - 41 |
ST05X1_TMJ |
Pt4w temp sensor, Straw station 5 |
ON |
42 - 43 |
ST05X1_TUS |
Pt4w temp sensor, Straw station 5 |
ON |
44 - 45 |
ST05X1_TUM |
Pt4w temp sensor, Straw station 5 |
ON |
46 - 47 |
ST05X1_TUJ |
Pt4w temp sensor, Straw station 5 |
ON |
48 |
ST05V1_H |
Humidity sensor, Straw station 5 |
ON |
49 |
ST05Y1_H |
Humidity sensor, Straw station 5 |
ON |
50 |
ST05X1_H |
Humidity sensor, Straw station 5 |
ON |
51 |
DL16_H |
Humidity sensor, Straw station 5 |
ON |
- ELMB #23
CANbus line #1 ADC mode: unipolar; ADC range: 100 mV. Place: bld.
888, W45 Rack. Status: installed.
ADC channels |
Dp name |
Comments |
Status |
0 - 1 |
W01_UP |
Pt4w temperature sensor |
ON |
2 - 3 |
W01_DOWN |
Pt4w temperature sensor |
ON |
4 - 5 |
W01_RIGHT |
Pt4w temperature sensor |
ON |
6 - 7 |
W01_LEFT |
Pt4w temperature sensor |
ON |
8 - 9 |
W02_UP |
Pt4w temperature sensor |
ON |
10 - 11 |
W02_DOWN |
Pt4w temperature sensor |
ON |
12 - 13 |
W02_LEFT |
Pt4w temperature sensor |
ON |
14 - 15 |
W02_RIGHT |
Pt4w temperature sensor |
ON |
16 - 17 |
W03_UP |
Pt4w temperature sensor |
ON |
18 - 19 |
W03_DOWN |
Pt4w temperature sensor |
ON |
20 - 21 |
W03_LEFT |
Pt4w temperature sensor |
ON |
22 - 23 |
W03_RIGHT |
Pt4w temperature sensor |
ON |
24 - 25 |
W04_UP |
Pt4w temperature sensor |
ON |
26 - 27 |
W04_DOWN |
Pt4w temperature sensor |
ON |
28 - 29 |
W04_LEFT |
Pt4w temperature sensor |
ON |
30 - 31 |
W04_RIGHT |
Pt4w temperature sensor |
ON |
32 - 33 |
W05_UP |
Pt4w temperature sensor |
ON |
34 - 35 |
W05_DOWN |
Pt4w temperature sensor |
ON |
36 - 37 |
W05_LEFT |
Pt4w temperature sensor |
ON |
38 - 39 |
W05_RIGHT |
Pt4w temperature sensor |
ON |
40 - 41 |
W06_UP |
Pt4w temperature sensor |
ON |
42 - 43 |
W06_DOWN |
Pt4w temperature sensor |
ON |
44 - 45 |
W06_LEFT |
Pt4w temperature sensor |
ON |
46 - 47 |
W06_RIGHT |
Pt4w temperature sensor |
ON |
48 - 49 |
W05_UP |
Pt4w temperature sensor |
ON |
50 - 51 |
T_Concrete |
Pt4w temperature sensor, T23 |
ON |
- ELMB #24
CANbus line #2 ADC mode: unipolar; ADC range: 100 mV. Place: bld. 888,
Gallery. Status: installed.
ADC channels |
Dp name |
Comments |
Status |
2 - 3 |
|
Pt4w temperature sensor |
OFF |
6 - 7 |
|
Pt4w temperature sensor |
OFF |
14 - 15 |
Env_T_Gallery |
Pt4w temperature sensor, T24 |
ON |
- ELMB #25
CANbus line #2 ADC mode: unipolar; ADC range: 100 mV. Place: bld.
888, RICH rack, Saleve side. Status: installed, but not connected to RICH, not
integrated in PVSS
- ELMB #29
CANbus line #2 ADC mode: unipolar; ADC range: 100 mV. Place: bld. 888,
Straw frame, Jura side. Comments: Status: installed.
ADC channels |
Dp name |
Comments |
Status |
0 - 1 |
ST02V2_TDS_up |
Pt4w temperature sensor |
ON |
2 - 3 |
ST02V2_TDM_up |
Pt4w temperature sensor |
ON |
4 - 5 |
ST02V2_TDJ_up |
Pt4w temperature sensor |
ON |
6 - 7 |
ST02V2_TMS_up |
Pt4w temperature sensor |
ON |
8 - 9 |
ST02V2_TMJ_up |
Pt4w temperature sensor |
ON |
10 - 11 |
ST02V2_TUS_up |
Pt4w temperature sensor |
ON |
12 - 13 |
ST02V2_TUM_up |
Pt4w temperature sensor |
ON |
14 - 15 |
ST02V2_TUJ_up |
Pt4w temperature sensor |
ON |
16 - 17 |
ST02Y2_TDS_up |
Pt4w temperature sensor |
ON |
18 - 19 |
ST02Y2_TDM_up |
Pt4w temperature sensor |
ON |
20 - 21 |
ST02Y2_TDJ_up |
Pt4w temperature sensor |
OFF |
22 - 23 |
ST02Y2_TMS_up |
Pt4w temperature sensor |
ON |
24 - 25 |
ST02Y2_TMJ_up |
Pt4w temperature sensor |
ON |
26 - 27 |
ST02Y2_TUS_up |
Pt4w temperature sensor |
ON |
28 - 29 |
ST02Y2_TUM_up |
Pt4w temperature sensor |
ON |
30 - 31 |
ST02Y2_TUJ_up |
Pt4w temperature sensor |
ON |
32 - 33 |
ST02X2_TDS_up |
Pt4w temperature sensor |
ON |
34 - 35 |
ST02X2_TDM_up |
Pt4w temperature sensor |
ON |
36 - 37 |
ST02X2_TDJ_up |
Pt4w temperature sensor |
ON |
38 - 39 |
ST02X2_TMS_up |
Pt4w temperature sensor |
ON |
40 - 41 |
ST02X2_TMJ_up |
Pt4w temperature sensor |
ON |
42 - 43 |
ST02X2_TUS_up |
Pt4w temperature sensor |
OFF |
44 - 45 |
ST02X2_TUM_up |
Pt4w temperature sensor |
ON |
46 - 47 |
ST02X2_TUJ_up |
Pt4w temperature sensor |
ON |
48 |
ST02V2_H_up |
Humidity sensor |
OFF |
49 |
ST02X2_H_up |
Humidity sensor |
OFF |
50 |
ST02Y2_H_up |
Humidity sensor |
OFF |
51 |
ST02X1_H_up |
Humidity sensor |
OFF |
52 |
ST02U1_H_up |
Humidity sensor |
OFF |
53 |
ST02Y1_H_up |
Humidity sensor |
OFF |
- ELMB #30
CANbus line #2 ADC mode: unipolar; ADC range: 100 mV. Place: bld. 888,
Straw frame, Jura side. Comments: Status: installed.
ADC channels |
Dp name |
Comments |
Status |
0 - 1 |
ST02V2_TDS_down |
Pt4w temperature sensor |
ON |
2 - 3 |
ST02V2_TDM_down |
Pt4w temperature sensor |
ON |
4 - 5 |
ST02V2_TDJ_down |
Pt4w temperature sensor |
OFF |
6 - 7 |
ST02V2_TMS_down |
Pt4w temperature sensor |
ON |
8 - 9 |
ST02V2_TMJ_down |
Pt4w temperature sensor |
ON |
10 - 11 |
ST02V2_TUS_down |
Pt4w temperature sensor |
ON |
12 - 13 |
ST02V2_TUM_down |
Pt4w temperature sensor |
ON |
14 - 15 |
ST02V2_TUJ_down |
Pt4w temperature sensor |
ON |
16 - 17 |
ST02Y2_TDS_down |
Pt4w temperature sensor |
ON |
18 - 19 |
ST02Y2_TDM_down |
Pt4w temperature sensor |
ON |
20 - 21 |
ST02Y2_TDJ_down |
Pt4w temperature sensor |
ON |
22 - 23 |
ST02Y2_TMS_down |
Pt4w temperature sensor |
OFF |
24 - 25 |
ST02Y2_TMJ_down |
Pt4w temperature sensor |
ON |
26 - 27 |
ST02Y2_TUS_down |
Pt4w temperature sensor |
OFF |
28 - 29 |
ST02Y2_TUM_down |
Pt4w temperature sensor |
ON |
30 - 31 |
ST02Y2_TUJ_down |
Pt4w temperature sensor |
ON |
32 - 33 |
ST02X2_TDS_down |
Pt4w temperature sensor |
ON |
34 - 35 |
ST02X2_TDM_down |
Pt4w temperature sensor |
ON |
36 - 37 |
ST02X2_TDJ_down |
Pt4w temperature sensor |
ON |
38 - 39 |
ST02X2_TMS_down |
Pt4w temperature sensor |
ON |
40 - 41 |
ST02X2_TMJ_down |
Pt4w temperature sensor |
ON |
42 - 43 |
ST02X2_TUS_down |
Pt4w temperature sensor |
ON |
44 - 45 |
ST02X2_TUM_down |
Pt4w temperature sensor |
ON |
46 - 47 |
ST02X2_TUJ_down |
Pt4w temperature sensor |
ON |
48 |
ST02V2_H_down |
Humidity sensor |
OFF |
49 |
ST02X2_H_down |
Humidity sensor |
OFF |
50 |
ST02Y2_H_down |
Humidity sensor |
OFF |
51 |
ST02X1_H_down |
Humidity sensor |
OFF |
52 |
ST02U1_H_down |
Humidity sensor |
OFF |
53 |
ST02Y1_H_down |
Humidity sensor |
OFF |
- ELMB #31
CANbus line #2 ADC mode: unipolar; ADC range: 100 mV. Place: bld. 888,
DC4, Jura side. Comments: Status: installed.
ADC channels |
Dp name |
Comments |
Status |
0 - 1 |
TT01 |
Pt4w temperature sensor |
ON |
2 - 3 |
TT02 |
Pt4w temperature sensor |
ON |
4 - 5 |
TT03 |
Pt4w temperature sensor |
ON |
6 - 7 |
TT04 |
Pt4w temperature sensor |
ON |
8 - 9 |
TT05 |
Pt4w temperature sensor |
ON |
10 - 11 |
TT06 |
Pt4w temperature sensor |
ON |
12 - 13 |
TT07 |
Pt4w temperature sensor |
ON |
14 - 15 |
TT08 |
Pt4w temperature sensor |
ON |
16 - 17 |
TT16 |
Pt4w temperature sensor |
ON |
18 - 19 |
TT09 |
Pt4w temperature sensor |
ON |
20 - 21 |
TT10 |
Pt4w temperature sensor |
ON |
22 - 23 |
TT11 |
Pt4w temperature sensor |
ON |
24 - 25 |
TT12 |
Pt4w temperature sensor |
ON |
26 - 27 |
TT13 |
Pt4w temperature sensor |
ON |
28 - 29 |
TT14 |
Pt4w temperature sensor |
ON |
30 - 31 |
TT15 |
Pt4w temperature sensor |
ON |
32 - 33 |
TT20 |
Pt4w temperature sensor |
ON |
34 - 35 |
TT21 |
Pt4w temperature sensor |
ON |
36 - 37 |
TT22 |
Pt4w temperature sensor |
ON |
38 - 39 |
TT23 |
Pt4w temperature sensor |
ON |
40 - 41 |
TT24 |
Pt4w temperature sensor |
ON |
42 - 43 |
TT25 |
Pt4w temperature sensor |
ON |
44 - 45 |
TT26 |
Pt4w temperature sensor |
ON |
46 - 47 |
TT27 |
Pt4w temperature sensor |
ON |
54 - 55 |
Mag_Sm1_T_Water_Return |
Pt4w temperature sensor |
ON |
56 - 57 |
Mag_Sm1_T_Water_Depart |
Pt4w temperature sensor |
ON |
58 - 59 |
T_Raw_Water_Return |
Pt4w temperature sensor |
ON |
60 - 61 |
T_Air_Duct_DC_MM |
Pt4w temperature sensor |
ON |
62 - 63 |
T_Raw_Water_Depart |
Pt4w temperature sensor |
ON |
- ELMB #32
CANbus
line #2 ADC mode: unipolar; ADC range: 100 mV. Place: bld. 888, W45
Rack, Saleve side.
Comments: ELMB used for monitoring of ECAL2 LV. Status: installed.
ADC channels |
Dp name |
Comments |
Status |
0 |
Ecal2_Lv_C0 |
ECAL2 LV distributor, column 0 |
ON |
1 |
Ecal2_Lv_C1 |
ECAL2 LV distributor, column 1 |
ON |
2 |
Ecal2_Lv_C2 |
ECAL2 LV distributor, column 2 |
ON |
3 |
Ecal2_Lv_C3 |
ECAL2 LV distributor, column 3 |
ON |
4 |
Ecal2_Lv_C4 |
ECAL2 LV distributor, column 4 |
ON |
5 |
Ecal2_Lv_C5 |
ECAL2 LV distributor, column 5 |
ON |
6 |
Ecal2_Lv_C6 |
ECAL2 LV distributor, column 6 |
ON |
7 |
Ecal2_Lv_C7 |
ECAL2 LV distributor, column 7 |
ON |
8 |
Ecal2_Lv_C8 |
ECAL2 LV distributor, column 8 |
ON |
9 |
Ecal2_Lv_C9 |
ECAL2 LV distributor, column 9 |
ON |
10 |
Ecal2_Lv_Col10 |
ECAL2 LV distributor, column 10 |
ON |
11 |
Ecal2_Lv_Col11 |
ECAL2 LV distributor, column 11 |
ON |
12 |
Ecal2_Lv_Col12 |
ECAL2 LV distributor, column 12 |
ON |
13 |
Ecal2_Lv_Col13 |
ECAL2 LV distributor, column 13 |
ON |
14 |
Ecal2_Lv_Col14 |
ECAL2 LV distributor, column 14 |
ON |
15 |
Ecal2_Lv_Col15 |
ECAL2 LV distributor, column15 |
ON |
27 |
Ecal2_Lv_for_LED_Neg_Top |
Low voltage of ECAL2 LED system |
ON |
28 |
Ecal2_Lv_for_LED_Neg_Bot |
Low voltage of ECAL2 LED system |
ON |
29 |
Ecal2_Lv_for_LED_Pos |
Low voltage of ECAL2 LED system |
ON |
30 |
Ecal2_Lv_for_Hv_2 |
Low voltage of ECAL2 for HV, "6 V" |
ON |
31 |
Ecal2_Lv_for_Hv_1 |
Low voltage of ECAL2 for HV, "6 V" |
ON |
32 |
Ecal2_Lv_Col16 |
ECAL2 LV distributor, column 16 |
ON |
33 |
Ecal2_Lv_Col17 |
ECAL2 LV distributor, column 17 |
ON |
34 |
Ecal2_Lv_C18 |
ECAL2 LV distributor, column 18 |
ON |
35 |
Ecal2_Lv_C19 |
ECAL2 LV distributor, column 19 |
ON |
36 |
Ecal2_Lv_C20 |
ECAL2 LV distributor, column 20 |
ON |
37 |
Ecal2_Lv_Col21 |
ECAL2 LV distributor, column 21 |
ON |
38 |
Ecal2_Lv_Col22 |
ECAL2 LV distributor, column 22 |
ON |
39 |
Ecal2_Lv_Col23 |
ECAL2 LV distributor, column 23 |
ON |
40 |
Ecal2_Lv_Col24 |
ECAL2 LV distributor, column 24 |
ON |
41 |
Ecal2_Lv_Col25 |
ECAL2 LV distributor, column 25 |
ON |
42 |
Ecal2_Lv_Col26 |
ECAL2 LV distributor, column 26 |
ON |
43 |
Ecal2_Lv_Col27 |
ECAL2 LV distributor, column 27 |
ON |
44 |
Ecal2_Lv_Col28 |
ECAL2 LV distributor, column 28 |
ON |
45 |
Ecal2_Lv_Col29 |
ECAL2 LV distributor, column 29 |
ON |
46 |
Ecal2_Lv_Col30 |
ECAL2 LV distributor, column 30 |
ON |
47 |
Ecal2_Lv_Col31 |
ECAL2 LV distributor, column 31 |
ON |
62 - 63 |
pt4w_elmb32_T63 |
Pt4w temperature sensor, T32 |
ON |
- ELMB #33
CANbus line #1 ADC mode: unipolar; ADC range: 100 mV. Place: bld. 888,
W45 rack, Saleve side. Comments:ELMB used for monitoring of ECAL2 LV. Status: installed.
ADC channels |
Dp name |
Comments |
Status |
32 |
Ecal2_Lv_Col48 |
ECAL2 LV distributor, column 48 |
ON |
33 |
Ecal2_Lv_Col49 |
ECAL2 LV distributor, column 49 |
ON |
34 |
Ecal2_Lv_Col50 |
ECAL2 LV distributor, column 50 |
ON |
35 |
Ecal2_Lv_Col51 |
ECAL2 LV distributor, column 51 |
ON |
36 |
Ecal2_Lv_Col48 |
ECAL2 LV distributor, column 52 |
ON |
37 |
Ecal2_Lv_Col53 |
ECAL2 LV distributor, column 53 |
ON |
38 |
Ecal2_Lv_Col54 |
ECAL2 LV distributor, column 54 |
ON |
39 |
Ecal2_Lv_Col55 |
ECAL2 LV distributor, column 55 |
ON |
40 |
Ecal2_Lv_Col56 |
ECAL2 LV distributor, column 56 |
ON |
41 |
Ecal2_Lv_Col57 |
ECAL2 LV distributor, column 57 |
ON |
42 |
Ecal2_Lv_Col58 |
ECAL2 LV distributor, column 58 |
ON |
43 |
Ecal2_Lv_Col59 |
ECAL2 LV distributor, column 59 |
ON |
44 |
Ecal2_Lv_Col60 |
ECAL2 LV distributor, column 60 |
ON |
45 |
Ecal2_Lv_Col61 |
ECAL2 LV distributor, column 61 |
ON |
46 |
Ecal2_Lv_Col62 |
ECAL2 LV distributor, column 62 |
ON |
47 |
Ecal2_Lv_Col48 |
ECAL2 LV distributor, column 63 |
ON |
32 |
Ecal2_Lv_Col48 |
ECAL2 LV distributor, column 48 |
ON |
33 |
Ecal2_Lv_Col49 |
ECAL2 LV distributor, column 49 |
ON |
34 |
Ecal2_Lv_Col50 |
ECAL2 LV distributor, column 50 |
ON |
35 |
Ecal2_Lv_Col51 |
ECAL2 LV distributor, column 51 |
ON |
36 |
Ecal2_Lv_Col52 |
ECAL2 LV distributor, column 52 |
ON |
37 |
Ecal2_Lv_Col53 |
ECAL2 LV distributor, column 53 |
ON |
38 |
Ecal2_Lv_Col54 |
ECAL2 LV distributor, column 54 |
ON |
39 |
Ecal2_Lv_Col55 |
ECAL2 LV distributor, column 55 |
ON |
40 |
Ecal2_Lv_Col56 |
ECAL2 LV distributor, column 56 |
ON |
41 |
Ecal2_Lv_Col57 |
ECAL2 LV distributor, column 57 |
ON |
42 |
Ecal2_Lv_Col58 |
ECAL2 LV distributor, column 58 |
ON |
43 |
Ecal2_Lv_Col59 |
ECAL2 LV distributor, column 59 |
ON |
44 |
Ecal2_Lv_Col60 |
ECAL2 LV distributor, column 60 |
ON |
45 |
Ecal2_Lv_Col61 |
ECAL2 LV distributor, column 61 |
ON |
46 |
Ecal2_Lv_Col62 |
ECAL2 LV distributor, column 62 |
ON |
47 |
Ecal2_Lv_Col63 |
ECAL2 LV distributor, column 63 |
ON |
- ELMB #34
CANbus line #1 ADC mode: unipolar; ADC range: 100 mV. Place: bld. 888,
W45 rack, Saleve side. Comments: ELMB prepared for monitoring of ECAL2 LV, but not connected to the detector. Status: installed.
- ELMB #35
CANbus line #1 ADC mode: unipolar; ADC range: 100 mV. Place: bld. 888,
W45 rack, Saleve side. Comments: ELMB prepared for monitoring of ECAL2 LV, but not connected to the detector. Status: installed.
Last update: August 9th, 2016
|
|